Hexagon Measurement      Systems

No matter what Hexagon measurement equipment or software you use, we want to hear your ideas and suggestions on how we can improve.

  • Your feedback and voting will benefit the majority of users.
  • You have 10 votes which you can allocate to your own or other users ideas.
  • We will focus on trying to realize the top voted items in our future products.

Thanks for your assistance in helping us shape the future.

+5

Reference Dimensions - No Tolerance & Secondary Reporting Options

Kyle Brummans hace 7 años en Metrology Software / PC-DMIS actualizado por AKASH hace 2 años 2

Two main attributes:

The ability to toggle tolerances off.  (Basic and Reference dimensions)  

The ability to have a second set of dimensional reporting options or several reporting profiles (Dimension Color Settings)

This helps with quality control.  Right now, we either 'open' tolerance or don't report.  There are instances where it is desirable to have values of basics and references with-out it hindering production mentality/reactions.  Often I also output XYZ coordinates for In-Car and zero tolerance them, as the T-value is what's important/desired.

+5

Increment Moves with Virtual Machine

davehocum hace 7 años en Metrology Software / PC-DMIS actualizado hace 5 años 3

Be able to use Increment Moves with Virtual Machine when executing paths and for collision detection.

+5

Provide tolerance zone visualization

Kesumm hace 7 años en Metrology Software / PC-DMIS actualizado por neil kay hace 6 años 0

Show tolerance zone visualization with DRF animation of GDT constraints

+5

PC-DMIS in one language, the report in another

Jörgen Andersson (vpt se) hace 8 años en Metrology Software / PC-DMIS actualizado por neil kay hace 6 años 0

I am running PC-DMIS with swedish language setting and thus, the report is also in swedish. However, it would be optimal if I could change the report template language to english - just for the report as we have customers all around the world.


In order to that today, I'd have to either run PC-DMIS in english (all the time) or change language, start PC-DMIS, open my program and print the report.


Would it be possible to add the possibility for the report templates to use a different language (other resources from the .dll's) template only for printing?

+5

Continuing development of Textonly Legacy Reporting

Simon Prodan hace 9 años en Metrology Software / PC-DMIS actualizado por neil kay hace 6 años 2

To have Legacy Textonly report supporting PC-DMIS Dimensions functionalities, or to have a similar report that is true text report and not just pictures that look like text.

+5

Auto circle

Krzysztof Malinowski hace 9 años en Metrology Software / PC-DMIS actualizado por neil kay hace 6 años 0

Add auto circle for star probe

+5
COMPLETADO

Datum definition allow multiple characters

Valentin Tarba hace 9 años en Metrology Software / PC-DMIS actualizado por Rob Jensen hace 6 años 14 1 duplicado

Allow to add multiple characters on a datum definition not only 2.

Image 22

+5
COMPLETADO

New point construction method: "high point"

dave amstutz (djams) hace 9 años en Metrology Software / PC-DMIS actualizado por Neil Ryan hace 2 años 2

Finding the "high point" of a feature or scan is a very common measurement task.


In pc-dmis, this is a complex task, that requires an understanding of array indexing, and significant manual entry into a generic feature. It is prone to error, is very time consuming, and for users never exposed to a programming language it's a complete non-starter.


A simple, interface-based point construction method could be added to address these issues. The new method should be based on the DMIS method "CONST/EXTREM". It is very complete, and since pc-dmis imports DMIS code - it only makes sense to have a 1 to1 mapping of the DMIS definition.

+5

Basic Dimension option

gsciola hace 9 años en Metrology Software / PC-DMIS actualizado por neil kay hace 6 años 0

How about a Basic Dimension option, in the Location Tool Box. That way the basic dimensions don't need to have tolerances.

+5

Angular Length Reporting

SalihaGungor hace 9 años en Metrology Software / PC-DMIS actualizado por neil kay hace 6 años 1

Image 12Maybe there is an easy way to report it ;if we have either a check box on the location (angle between) dialog window like "from formula" or an easier assignment set will be great.

Here is the code in order to report it :


Image 14

CIR1 =FEAT/CIRCLE,CARTESIAN,OUT,LEAST_SQR
THEO/<98.516,-145.613,-2.547>,<0,0,1>,506.5,0
ACTL/<98.516,-145.613,-2.547>,<0,0,1>,506.5,0
MEAS/CIRCLE,4,ZPLUS
HIT/BASIC,NORMAL,<51.696,103.272,-2.55>,<-0.1848771,0.9827616,0>,<51.696,103.272,-2.55>,USE THEO=YES HIT/BASIC,NORMAL,<99.287,107.636,-1.309>,<0.0030416,0.9999954,0>,<99.287,107.636,-1.309>,USE THEO=YES
HIT/BASIC,NORMAL,<166.819,98.253,-3.669>,<0.2697032,0.9629435,0>,<166.819,98.253,-3.669>,USE THEO=YES
HIT/BASIC,NORMAL,<227.698,72.212,-2.661>,<0.5100969,0.8601169,0>,<227.698,72.212,-2.661>,USE THEO=YES
ENDMEAS/
LIN1 =FEAT/LINE,CARTESIAN,UNBOUNDED
THEO/<31.03,97.103,-1.56>,<-0.7071068,-0.7071068,0>
ACTL/<31.03,97.103,-1.56>,<-0.7071068,-0.7071068,0>
MEAS/LINE,2,ZPLUS
HIT/BASIC,NORMAL,<31.03,97.103,1.086>,<-0.7071068,0.7071068,0>,<31.03,97.103,1.086>,USE THEO=YES
HIT/BASIC,NORMAL,<-13.352,52.72,-4.207>,<-0.7071068,0.7071068,0>,<-13.352,52.72,-4.207>,USE THEO=YES
ENDMEAS/
LIN2 =FEAT/LINE,CARTESIAN,UNBOUNDED
THEO/<239,62.251,-5.264>,<0,-1,0>
ACTL/<239,62.251,-5.264>,<0,-1,0>
MEAS/LINE,2,ZPLUS
HIT/BASIC,NORMAL,<239,62.251,-5.016>,<1,0,0>,<239,62.251,-5.016>,USE THEO=YES
HIT/BASIC,NORMAL,<239,53.115,-5.512>,<1,0,0>,<239,53.115,-5.512>,USE THEO=YES
ENDMEAS/
PNT1 =FEAT/POINT,CARTESIAN,NO
THEO/<239,65.1,-5.264>,<0,-1,0>
ACTL/<239,65.1,-5.264>,<0,-1,0>
CONSTR/POINT,PIERCE,CIR1,LIN2
PNT2 =FEAT/POINT,CARTESIAN,NO
THEO/<32.922,98.995,-1.56>,<-0.7071068,-0.7071068,0>
ACTL/<32.922,98.995,-1.56>,<-0.7071068,-0.7071068,0>
CONSTR/POINT,PIERCE,CIR1,LIN1
LIN3 =FEAT/LINE,CARTESIAN,UNBOUNDED,NO
THEO/<98.516,-145.613,-2.547>,<-0.259008,0.9658673,0.0038973>
ACTL/<98.516,-145.613,-2.547>,<-0.259008,0.9658673,0.0038973>
CONSTR/LINE,BF,3D,CIR1,PNT2,,
OUTLIER_REMOVAL/OFF,3
FILTER/OFF,WAVELENGTH=0
LIN4 =FEAT/LINE,CARTESIAN,UNBOUNDED,NO
THEO/<98.516,-145.613,-2.547>,<0.5546916,0.8319869,-0.0107257>
ACTL/<98.516,-145.613,-2.547>,<0.5546916,0.8319869,-0.0107257>
CONSTR/LINE,BF,3D,CIR1,PNT1,,
OUTLIER_REMOVAL/OFF,3
FILTER/OFF,WAVELENGTH=0
DIM ANGL1= 2D ANGLE FROM LINE LIN4 TO LINE LIN3 ,$
GRAPH=OFF TEXT=OFF MULT=10.00 OUTPUT=BOTH
AX MEAS NOMINAL +TOL -TOL DEV OUTTOL
A 48.703 48.703 0.010 0.010 0.000 0.000 ----#----
ASSIGN/V1=((2*3.14*CIR1.R.MEAS)/360)*ANGL1.A.MEAS
F1 =GENERIC/NONE,DEPENDENT,CARTESIAN,OUT,$
NOM/XYZ,<0,0,0>,$ MEAS/XYZ,<0,0,0>,$
NOM/IJK,<0,0,1>,$
MEAS/IJK,<0,0,1>,$
RADIUS/0,0,$
ANGLE/215.5,V1,$
DISTANCE/0,0
DIM LOC1= LOCATION OF PLANE F1 UNITS=MM ,$
GRAPH=OFF TEXT=OFF MULT=10.00 OUTPUT=BOTH HALF ANGLE=NO
AX MEAS NOMINAL +TOL -TOL DEV OUTTOL
A 215.160 215.500 0.500 0.500 -0.340 0.000 -#-------
END OF DIMENSION LOC1