+4

2020 R2 Profile - CAD graphic analysis improvement

Jacob Cheverie 4 года назад в Metrology Software / PC-DMIS обновлен Neil Challinor 2 года назад 3

In 2020 R2, the calculations/interpretations of profile tolerances have changed. The idea now, for a unilateral tolerance, is that the tolerance zone is offset from the nominal surface an appropriate amount and the zone is expanded/contracted until the surface is "snug". 

I created a simple example. Offline, I created an external cylinder. I requested the profile of the surface with [.xxx U 0]. I turned on CAD graphic analysis and the tolerance zone is still centered on nominal. What's worse is that the deviation shows toward the wrong extreme of the zone.

Can this feature be improved so that the graphic analysis of the zone is shifted and the deviation can be shown properly?

Image 429

As you can see, at [.xxx U 0], the outer edge would be at the extreme outer edge of the zone. PC-DMIS is showing it at the inner edge.

+2

This has been added for 2023.1.

@Aaron Baldauf


Not quite, what we did in 2023.1 was to scale the nominal as well as the tolerance lines.  Previously, the tolerance lines scaled but the nominal did not, making it difficult to see the offset.

Сервис поддержки клиентов работает на платформе UserEcho